Yıl: 2018 Cilt: 7 Sayı: 3 Sayfa Aralığı: 503 - 506 Metin Dili: İngilizce DOI: 10.5455/medscience.2018.07.8805 İndeks Tarihi: 19-09-2019

Evaluation of BT uricell1280 automated urine sediment analyzer performance

Öz:
Fully automated urine analyzers are generally preferred in laboratories with high throughput. When these systems are used; especially the sediment analysis should be made by a well-skilled technician before reporting to discriminate well the interferences (like yeast or crystals) from erythrocytes and leukocytes. The purpose of the present study is to detect the correlation of erythrocyte and leukocyte numbers without any corrections in sediment field images of BT Uricell 1280 (BT products, İzmir, Turkey) fully automated urine analyzer with manual microscopy reports. A total of 528 first morning urine samples were studied by both BT Uricell 1280 and manual microscopy concurrently. The degree of concordance (Kappa coefficient) were evaluated. The sensitivity, specificity and positive and negative predictive value for the BT Uricell 1280 compared to manual microscopic examination were assessed. Precision and carry over studies were also performed. The degree of concordance of erythrocyte and leukocyte counts in microscopy of 528 urine specimens with manual microscopy was found to be 0.42 and 0.52 respectively (kappa coefficient). The sensitivity and specificity values of RBC and WBC were calculated for BT Uricell 1280 as 62.6%, 91.1%, 77.6%, 88.8%, respectively. The results of carry over analysis for both RBC and WBC were 0 %. The sediment microscopy analysis of erythrocytes and leukocytes with BT Uricell 1280 without any corrections presented moderate correlation with manual microscopy. A better correlation can be achieved by a more detailed examination of field images of the analyzer.
Anahtar Kelime:

Belge Türü: Makale Makale Türü: Araştırma Makalesi Erişim Türü: Erişime Açık
  • Gadeholt H. Quantitative estimation of urinary sediment, with special regard to sources of error. Br Med J. 1964;1:1547-9.
  • Winkel P, Statland BE, Jorgensen K. Urine microscopy, an ill-defined method, examined by a multifactorial technique. Clin Chem. 1974;20:436-9.
  • European Confederation of Laboratory Medicine. European urinalysis guidelines. Scand J Clin Lab Invest Suppl. 2000;231:1-86.
  • Clemens AH, Hurtle RL. Automatic system for urine analysis. I. System design and development. Clinl Chem. 1972;18:789-93.
  • Chien TI, Kao JT, Liu HL, et al. Urine sediment examination: a comparison of automated urinalysis systems and manual microscopy. Clin Chim Acta. 2007;384:28-34.
  • Hyodo T, Kumano K, Haga M, et al. Detection of glomerular and nonglomerular red blood cells by automated urinary sediment analyzer. Nihon Jinzo Gakkai Shi. 1995;37:35-43.
  • Deindoerfer FH, Gangwer JR, Laird CW, et al. The Yellow IRIS” urinalysis workstation--the first commercial application of automated intelligent microscopy. Clin Chem. 1985;31:1491-9.
  • Okada H, Sakai Y, Kawabata G, et al. Automated urinalysis. Evaluation of the Sysmex UF-50. Am J Clin Pathol. 2001;115:605-10.
  • Elin RJ, Hosseini JM, Kestner J, et al. Comparison of automated and manual methods for urinalysis. Am J Clin Pathol. 1986;86:731-7.
  • Lamchiagdhase P, Preechaborisutkul K, Lomsomboon P, et al. Urine sediment examination: a comparison between the manual method and the iQ200 automated urine microscopy analyzer. Clin Chim Acta. 2005;358:167-74.
  • Fenili D, Pirovano B. The automation of sediment urinalysis using a new urine flow cytometer (UF-100). Clin Chem Lab Med. 1998;36:909-17.
  • Ma J, Wang C, Yue J, et al. Clinical laboratory urine analysis: comparison of the UriSed automated microscopic analyzer and the manual microscopy. Clin Lab. 2013;59:1297-303.
  • Yuksel H, Kilic E, Ekinci A, et al. Comparison of fully automated urine sediment analyzers H800-FUS100 and LabUMat-UriSed with manual microscopy. J Clin Lab Anal. 2013;27:312-6.
  • Bakan E, Ozturk N, Baygutalp NK, et al. Comparison of Cobas 6500 and Iris IQ200 fully-automated urine analyzers to manual urine microscopy. Biochem Med. 2016;26:365-75.
  • Shayanfar N, Tobler U, von Eckardstein A, et al. Automated urinalysis: first experiences and a comparison between the Iris iQ200 urine microscopy system, the Sysmex UF-100 flow cytometer and manual microscopic particle counting. Clin Chem Lab Med. 2007;45:1251-6.
  • . Akin OK, Serdar MA, Cizmeci Z, et al. Comparison of LabUMat-withUriSed and iQ200 fully automatic urine sediment analysers with manual urine analysis. Biotechnol Appl Biochem. 2009;53:139-44.
  • Dewulf G, Harrois D, Mazars E, et al. Evaluation of the performances of the iQ((R))200 ELITE automated urine microscopy analyser and comparison with manual microscopy method. Pathol Biol (Paris). 2011;59:264-8.
  • Linko S, Kouri TT, Toivonen E, et al. Analytical performance of the Iris iQ200 automated urine microscopy analyzer. Clin Chim Acta. 2006;372:5464.
  • Wah DT, Wises PK, Butch AW. Analytic performance of the iQ200 automated urine microscopy analyzer and comparison with manual counts using FuchsRosenthal cell chambers. Am J Clin Pathol. 2005;123:290-6.
APA ERCAN KARADAĞ M, FIRAT OĞUZ E (2018). Evaluation of BT uricell1280 automated urine sediment analyzer performance. , 503 - 506. 10.5455/medscience.2018.07.8805
Chicago ERCAN KARADAĞ MÜJGAN,FIRAT OĞUZ Esra Evaluation of BT uricell1280 automated urine sediment analyzer performance. (2018): 503 - 506. 10.5455/medscience.2018.07.8805
MLA ERCAN KARADAĞ MÜJGAN,FIRAT OĞUZ Esra Evaluation of BT uricell1280 automated urine sediment analyzer performance. , 2018, ss.503 - 506. 10.5455/medscience.2018.07.8805
AMA ERCAN KARADAĞ M,FIRAT OĞUZ E Evaluation of BT uricell1280 automated urine sediment analyzer performance. . 2018; 503 - 506. 10.5455/medscience.2018.07.8805
Vancouver ERCAN KARADAĞ M,FIRAT OĞUZ E Evaluation of BT uricell1280 automated urine sediment analyzer performance. . 2018; 503 - 506. 10.5455/medscience.2018.07.8805
IEEE ERCAN KARADAĞ M,FIRAT OĞUZ E "Evaluation of BT uricell1280 automated urine sediment analyzer performance." , ss.503 - 506, 2018. 10.5455/medscience.2018.07.8805
ISNAD ERCAN KARADAĞ, MÜJGAN - FIRAT OĞUZ, Esra. "Evaluation of BT uricell1280 automated urine sediment analyzer performance". (2018), 503-506. https://doi.org/10.5455/medscience.2018.07.8805
APA ERCAN KARADAĞ M, FIRAT OĞUZ E (2018). Evaluation of BT uricell1280 automated urine sediment analyzer performance. Medicine Science, 7(3), 503 - 506. 10.5455/medscience.2018.07.8805
Chicago ERCAN KARADAĞ MÜJGAN,FIRAT OĞUZ Esra Evaluation of BT uricell1280 automated urine sediment analyzer performance. Medicine Science 7, no.3 (2018): 503 - 506. 10.5455/medscience.2018.07.8805
MLA ERCAN KARADAĞ MÜJGAN,FIRAT OĞUZ Esra Evaluation of BT uricell1280 automated urine sediment analyzer performance. Medicine Science, vol.7, no.3, 2018, ss.503 - 506. 10.5455/medscience.2018.07.8805
AMA ERCAN KARADAĞ M,FIRAT OĞUZ E Evaluation of BT uricell1280 automated urine sediment analyzer performance. Medicine Science. 2018; 7(3): 503 - 506. 10.5455/medscience.2018.07.8805
Vancouver ERCAN KARADAĞ M,FIRAT OĞUZ E Evaluation of BT uricell1280 automated urine sediment analyzer performance. Medicine Science. 2018; 7(3): 503 - 506. 10.5455/medscience.2018.07.8805
IEEE ERCAN KARADAĞ M,FIRAT OĞUZ E "Evaluation of BT uricell1280 automated urine sediment analyzer performance." Medicine Science, 7, ss.503 - 506, 2018. 10.5455/medscience.2018.07.8805
ISNAD ERCAN KARADAĞ, MÜJGAN - FIRAT OĞUZ, Esra. "Evaluation of BT uricell1280 automated urine sediment analyzer performance". Medicine Science 7/3 (2018), 503-506. https://doi.org/10.5455/medscience.2018.07.8805