Evaluation of PET Scanner Performance in PET/MR and PET/CT Systems: NEMA Tests

Öz:
Objective: The aim of the present study was to compare the performance of positron emission tomography (PET) component of PET/computed tomography (CT) with new emerging PET/magnetic resonance (MR) of the same vendor. Methods: According to National Electrical Manufacturers Association NU2-07, five separate experimental tests were performed to evaluate the performance of PET scanner of General Electric GE company; SIGNATM model PET/MR and GE Discovery 710 model PET/CT. The main investigated aspects were spatial resolution, sensitivity, scatter fraction, count rate performance, image quality, count loss and random events correction accuracy. Results: The findings of this study demonstrated superior sensitivity (~ 4 folds) of PET scanner in PET/MR compared to PET/ CT system. Image quality test exhibited higher contrast in PET/MR (~ 9%) compared with PET/CT. The scatter fraction of PET/ MR was 43.4% at noise equivalent count rate (NECR) peak of 218 kcps and the corresponding activity concentration was 17.7 kBq/cc. Whereas the scatter fraction of PET/CT was found as 39.2% at NECR peak of 72 kcps and activity concentration of 24.3 kBq/cc. The percentage error of the random event correction accuracy was 3.4% and 3.1% in PET/MR and PET/CT, respectively. Conclusion: It was concluded that PET/MR system is about 4 times more sensitive than PET/CT, and the contrast of hot lesions in PET/MR was ~ 9% higher than PET/CT. These outcomes also emphasize the possibility to achieve excellent clinical PET images with low administered dose and/or a short acquisition time in PET/MR.
Anahtar Kelime:

Konular: Radyoloji, Nükleer Tıp, Tıbbi Görüntüleme

PET/MR ve PET/BT Sistemlerinde PET Tarayıcının Performanslarının Değerlendirilmesi: NEMA Testleri

Öz:
Amaç: Bu çalışmada son yıllarda nükleer tıpta rutin uygulamaya giren pozitron emisyon tomografisi/manyetik rezonans (PET/ MR) görüntüleme sistemi ile PET/bilgisayarlı tomografi (BT) görüntüleme sistemlerinde ‘National Electrical Manufacturers Association’ (NEMA) testlerinin yapılması ve sonuçların karşılaştırılması amaçlandı. Yöntem: NEMA NU2-07 testlerinden uzaysal ayırma gücü, hassasiyet, saçılma fraksiyonu kayıp sayımlar ve rastgele olay ölçümleri, görüntü kalitesi, sayım kayıpları rastgele olay düzeltme doğruluğu olmak üzere beş ayrı deneysel çalışma yapıldı. Bulgular: Sensitivite testi sonuçları değerlendirildiğinde PET/MR sisteminin PET/BT sistemine göre yaklaşık 4 kat daha üstün olduğu bulundu. Görüntü kalitesi test sonuçlarından sıcak lezyon kontrastının PET/MR’da %9 daha yüksek olduğu bulundu. PET/MR scatter oranı gürültü eşdeğeri sayım hızı (NECR) 218 kcts ve aktivite konsantrasyonunun 17,7 kcts/cc olduğu durumda %43,4 bulundu. Buna karşılık PET/BT scatter oranı NECR 72 kcts ve aktivite konsantrasyonunun 24,3 kcts/cc olduğu durumda %39,2 bulundu. Tesadüfi hataları düzeltme doğruluğu PET/MR’da 3,4 ve PET/BT’de 3,1 bulundu. Sonuç: PET/MR’da sensitivite değerinin PET/BT’ye göre yaklaşık 4 kat daha yüksek olduğu, PET/MR’da sıcak lezyon kontrastının PET/BT’ye göre %9 daha yüksek olduğu ayrıca PET/MR’da PET görüntülerinin daha az aktivite miktarı/veya daha kısa çekim süresinde elde edilebileceği sonucuna varıldı.
Anahtar Kelime:

Konular: Radyoloji, Nükleer Tıp, Tıbbi Görüntüleme
Belge Türü: Makale Makale Türü: Araştırma Makalesi Erişim Türü: Erişime Açık
  • Torigian DA, Habib Zaidi H, Kwee TC, Saboury B, Udupa JK, Cho ZH, Alavi A. PET/MR Imaging: Technical Aspects and Potential Clinical Applications. Radiology 2013;267:26-44.
  • Roncani E, Cherry SR. Application of silicon photomultipliers to positron emission tomography. Ann Biomed Eng 2011;39:13581377.
  • Britvich I, Johnson I, Renker D, Stoykov A, Lorenz E. Characterization of Geiger-mode avalanche photodiodes for medical imaging applications. Nucl Instrum Methods Phys Res A 2007;43:2044-2048.
  • Schaart DR, Seifert S, Vinke R, van Dam HT, Dendooven P, Löhner H, Beekman FJ. LaBr(3):Ce and SiPMs for time-of-flight PET: Achieving 100 ps coincidence resolving time. Phys Med Biol 2011;56:4135-4145.
  • Martinez-Möller A, Nekolla SG. Attenuation correction for PET/MR: problems, novel approaches and practical solutions. Z Med Phys 2012;22:299-310.
  • Martinez-Möller A, Souvatzoglou M, Delso G, Bundschuh RA, Chefd’hotel C, Ziegler SI, Navab N, Schwaiger M, Nekolla SG. Tissue classification as a potential approach for attenuation correction in whole-body PET/MRI: evaluation with PET/CT data. J Nucl Med 2009;50:520-526.
  • National Electrical Manufacturers Association. “NEMA Standards Publication NU 2-2007, Performance measurements of positron emission tomographs”. Rosslyn VA 2007;26-33.
  • Sorenson JA, Phelp ME. Physics in Nuclear Medicine. Grune and Stration, Inc, 2003;227-251.
  • National Electrical Manufacturers Association. Performance measurements of position emission tomographs NEMA Standards Publication NU 2-2007. 2007.
  • . Yester M, Senan RA, White S. NEMA testing of GE Discovery 710 PET scanner compared to a simplified protocol for routine testing of PET scanners. J Nucl Med 2014;55:2157.
  • Grant AM, Deller TW, Khalighi MM, Maramraju SH, Delso G, Levin CS. NEMA NU 2-2012 performance studies for the SiPM-based ToF-PET component of the GE SIGNA PET/MR system. Med Phys 2016;43:2334.
  • Ziegler S, Jakoby BW, Braun H, Paulus DH, Quick HH. NEMA image quality phantom measurements and attenuation correction in integrated PET/MR hybrid imaging. EJNMMI Phys 2015;2:18.
  • Drzezga A, Souvatzoglou M, Eiber M, Beer AJ, Fürst S, MartinezMöller A, Nekolla SG, Ziegler S, Ganter C, Rummeny EJ, Schwaiger M. First clinical experience with integrated whole-body PET/MR: comparison to PET/CT in patients with oncologic diagnoses. J Nucl Med 2012;53:845-855.
  • Wiesmüller M, Quick HH, Navalpakkam B, Lell MM, Uder M, Ritt P, Schmidt D, Beck M, Kuwert T, Von Gall CC. Comparison of lesion detection and quantitation of tracer uptake between PET from a simultaneously acquiring whole-body PET/MR hybrid scanner and PET from PET/CT. Eur J Nucl Med Mol Imaging 2013;40:12-21.
  • Oehmigen M, Ziegler S, Jakoby BW, Georgi JC, Paulus DH, Quick HH. Radiotracer Dose Reduction in Integrated PET/MR: Implications from National Electrical Manufacturers Association Phantom Studies. J Nucl Med 2014;55:1361-1367.
  • Martinez-Möller A, Souvatzoglou M, Delso G, Bundschuh RA, Chefd’hotel C, Ziegler SI, Navab N, Schwaiger M, Nekolla SG. Tissue classification as a potential approach for attenuation correction in whole-body PET/MRI. J Nucl Med 2009;50:520-526.
  • Delso G, Fürst S, Jakoby B, Ladebeck R, Ganter C, Nekolla SG, Schwaiger M, Ziegler SI. Performance measurements of the Siemens mMR integrated whole-body PET/MR scanner. J Nucl Med 2011;52:1914-1922.
  • Karlberg AM, Sæther O, Eikenes L, Goa PE. Quantitative comparison of PET Performance-Siemens Biograph mCT and mMR. EJNMMI Phys 2016;3:5.
APA DEMİR M, TOKLU T, ABUQBEITAH M, ÇETİN H, SEZGİN H, Yeyin N, SÖNMEZOĞLU K (2018). Evaluation of PET Scanner Performance in PET/MR and PET/CT Systems: NEMA Tests. , 10 - 18. 10.4274/mirt.97659
Chicago DEMİR MUSTAFA,TOKLU Türkay,ABUQBEITAH Mohammad,ÇETİN HÜSEYİN,SEZGİN H. Sezer,Yeyin Nami,SÖNMEZOĞLU Kerim Evaluation of PET Scanner Performance in PET/MR and PET/CT Systems: NEMA Tests. (2018): 10 - 18. 10.4274/mirt.97659
MLA DEMİR MUSTAFA,TOKLU Türkay,ABUQBEITAH Mohammad,ÇETİN HÜSEYİN,SEZGİN H. Sezer,Yeyin Nami,SÖNMEZOĞLU Kerim Evaluation of PET Scanner Performance in PET/MR and PET/CT Systems: NEMA Tests. , 2018, ss.10 - 18. 10.4274/mirt.97659
AMA DEMİR M,TOKLU T,ABUQBEITAH M,ÇETİN H,SEZGİN H,Yeyin N,SÖNMEZOĞLU K Evaluation of PET Scanner Performance in PET/MR and PET/CT Systems: NEMA Tests. . 2018; 10 - 18. 10.4274/mirt.97659
Vancouver DEMİR M,TOKLU T,ABUQBEITAH M,ÇETİN H,SEZGİN H,Yeyin N,SÖNMEZOĞLU K Evaluation of PET Scanner Performance in PET/MR and PET/CT Systems: NEMA Tests. . 2018; 10 - 18. 10.4274/mirt.97659
IEEE DEMİR M,TOKLU T,ABUQBEITAH M,ÇETİN H,SEZGİN H,Yeyin N,SÖNMEZOĞLU K "Evaluation of PET Scanner Performance in PET/MR and PET/CT Systems: NEMA Tests." , ss.10 - 18, 2018. 10.4274/mirt.97659
ISNAD DEMİR, MUSTAFA vd. "Evaluation of PET Scanner Performance in PET/MR and PET/CT Systems: NEMA Tests". (2018), 10-18. https://doi.org/10.4274/mirt.97659
APA DEMİR M, TOKLU T, ABUQBEITAH M, ÇETİN H, SEZGİN H, Yeyin N, SÖNMEZOĞLU K (2018). Evaluation of PET Scanner Performance in PET/MR and PET/CT Systems: NEMA Tests. Molecular Imaging and Radionuclide Therapy, 27(1), 10 - 18. 10.4274/mirt.97659
Chicago DEMİR MUSTAFA,TOKLU Türkay,ABUQBEITAH Mohammad,ÇETİN HÜSEYİN,SEZGİN H. Sezer,Yeyin Nami,SÖNMEZOĞLU Kerim Evaluation of PET Scanner Performance in PET/MR and PET/CT Systems: NEMA Tests. Molecular Imaging and Radionuclide Therapy 27, no.1 (2018): 10 - 18. 10.4274/mirt.97659
MLA DEMİR MUSTAFA,TOKLU Türkay,ABUQBEITAH Mohammad,ÇETİN HÜSEYİN,SEZGİN H. Sezer,Yeyin Nami,SÖNMEZOĞLU Kerim Evaluation of PET Scanner Performance in PET/MR and PET/CT Systems: NEMA Tests. Molecular Imaging and Radionuclide Therapy, vol.27, no.1, 2018, ss.10 - 18. 10.4274/mirt.97659
AMA DEMİR M,TOKLU T,ABUQBEITAH M,ÇETİN H,SEZGİN H,Yeyin N,SÖNMEZOĞLU K Evaluation of PET Scanner Performance in PET/MR and PET/CT Systems: NEMA Tests. Molecular Imaging and Radionuclide Therapy. 2018; 27(1): 10 - 18. 10.4274/mirt.97659
Vancouver DEMİR M,TOKLU T,ABUQBEITAH M,ÇETİN H,SEZGİN H,Yeyin N,SÖNMEZOĞLU K Evaluation of PET Scanner Performance in PET/MR and PET/CT Systems: NEMA Tests. Molecular Imaging and Radionuclide Therapy. 2018; 27(1): 10 - 18. 10.4274/mirt.97659
IEEE DEMİR M,TOKLU T,ABUQBEITAH M,ÇETİN H,SEZGİN H,Yeyin N,SÖNMEZOĞLU K "Evaluation of PET Scanner Performance in PET/MR and PET/CT Systems: NEMA Tests." Molecular Imaging and Radionuclide Therapy, 27, ss.10 - 18, 2018. 10.4274/mirt.97659
ISNAD DEMİR, MUSTAFA vd. "Evaluation of PET Scanner Performance in PET/MR and PET/CT Systems: NEMA Tests". Molecular Imaging and Radionuclide Therapy 27/1 (2018), 10-18. https://doi.org/10.4274/mirt.97659