Yıl: 2020 Cilt: 35 Sayı: 3 Sayfa Aralığı: 699 - 709 Metin Dili: Türkçe İndeks Tarihi: 07-05-2021

Evaluation of CNT/MOSFET Based Active Electrode with ECG, EMG, and EEG Signals

Öz:
In this study, the electrical circuit equivalents of the previously designed novel CNT/MOSFET basedactive electrode and commonly used wet electrode are simulated by applying different biopotentials.Responses of the electrode models are compared within the certain skin surface circumstances. Fouriertransforms and total harmonic distortion calculations of obtained biopotentials are examined to assess theelectrodes. The simulation results demonstrate that the CNT / MOSFET based electrode predominates thewet electrode and can measure biopotentials with high quality.
Anahtar Kelime:

KNT/MOSFET Tabanlı Aktif Elektrotun EKG, EMG, EEG İşaretleri ile Değerlendirilmesi

Öz:
Bu çalışmada, önceden tasarlanmış olan özgün KNT/MOSFET tabanlı aktif elektrotun ve yaygın olarak kullanılan ıslak elektrotun elektrik devresi eşdeğerlerine, farklı biyopotansiyeller uygulanarak benzetimleri gerçekleştirilmiştir. Elektrotların giriş sinyallerine verdiği yanıtlar, belirli cilt yüzeyi koşullarında karşılaştırılmıştır. Elektrotları değerlendirmek için elde edilen biyopotansiyellerin Fourier dönüşümleri ve toplam harmonik bozulmaları incelenmiştir. Simülasyon sonuçları, KNT/MOSFET tabanlı elektrotun ıslak elektrottan daha iyi sonuçlar verdiğini ve biyopotansiyelleri yüksek kalitede ölçebileceğini göstermektedir
Anahtar Kelime:

Belge Türü: Makale Makale Türü: Araştırma Makalesi Erişim Türü: Erişime Açık
  • 1. Grimnes, S., Martinsen, O.G., 2015.. Bioimpedance And Bioelectricity Basics, Academic Press, 3rd Edt.
  • 2. Plonsey, R., Barr, R.C., 2007. Bioelectricity: A Quantitative Approach, Springer US, 3rd Edt.
  • 3. Webster, J.G., Ed., 2006. Encyclopedia of Medical Devices and Instrumentation, Hoboken, NJ., USA: John Wiley & Sons, Inc.
  • 4. Neuman M., 2010. Biopotential Electrodes in Medical Instrumentation Application and Design, Fourth Edi., Webster J., Ed. Wiley, 189–240.
  • 5. Valchinov, E.S., Pallikarakis, N.E., 2004. An Active Electrode for Biopotential Recording from Small Localized Bio-sources, Biomed. Eng. Online, 3(1), 25.
  • 6. Yoo, J., Yan, L., Lee, S., Kim, H., Kim, B., Yoo, H.J., 2009. An Attachable ECG Sensor Bandage with Planar-fashionable Circuit Board, Proc.-Int. Symp. Wearable Comput. ISWC, 145–146.
  • 7. Wang, L.F., Liu, J.Q., Yang, B., Yang, C.S., 2012. PDMS-Based Low Cost Flexible Dry Electrode for Long-Term EEG Measurement, IEEE Sens. J., 12(9), 2898–2904.
  • 8. Griss, P., Tolvanen, H.K., Meriläinen, P., Stemme G., 2002. Characterization of Micromachined Spiked Biopotential Electrodes, IEEE Trans. Biomed. Eng., 49(6), 597–604.
  • 9. Mestrovic, M.A., Helmer, R.J.N., Kyratzis, L., Kumar, D., 2007. Preliminary Study of Dry Knitted Fabric Electrodes for Physiological Monitoring, Proc. 2007 Int. Conf. Intell. Sensors, Sens. Networks Inf. Process. ISSNIP, 601–606.
  • 10. Gargiulo, G., Bifulco, P., Calvo, R.A., Cesarelli, M., Jin, C., Van Schaik, A., 2008. Mobile Biomedical Sensing with Dry Electrodes, ISSNIP 2008-Proc. 2008 Int. Conf. Intell. Sensors, Sens. Networks Inf. Process., 261–266.
  • 11. Giulio, R., Stephen, D., Esteve, F., Josep, M.P., Chris, R., Ernest, M., Ravi, S., Carles, G., 2006. A Dry Electrophysiology Electrode Using CNT Arrays, Sensors Actuators A Phys., 132(1), 34–41.
  • 12. Abu-Saude, M.J., Morshed, B.I., 2015. Patterned Vertical Carbon Nanotube Dry Electrodes for Impedimetric Sensing and Stimulation, IEEE Sens. J., 15(10), 5851-5858.
  • 13. Abu-Saude, M., Consul-Pacareu, S., Morshed, B.I., 2015. Feasibility of Patterned Vertical CNT for Dry Electrode Sensing of Physiological Parameters, 2015 IEEE Topical Conference on Biomedical WirelessTechnologies, Networks, and Sensing Systems (BioWireleSS), San Diego, CA, 1-4.
  • 14. Istanbullu, M., Avci, M., 2019. Novel Multiwalled Carbon Nanotube (MWCNT) Modified Metal Oxide Semiconductor Field Effect Transistor (MOSFET) Based Electrode for Electrophysiological Measurements on Human Skin, Instrum. Sci. Technol., 47(5), 545–563.
  • 15. Liu, Y.L., Lin, D.L., Lin, Y.D., 2004. Spice Model for Computer-Aided Design of Biopotential Amplifier, Biomed. Eng. Appl. Basis Commun., 16(3), 151–156.
APA İstanbullu M, AVCI M (2020). Evaluation of CNT/MOSFET Based Active Electrode with ECG, EMG, and EEG Signals. , 699 - 709.
Chicago İstanbullu Mustafa,AVCI MUTLU Evaluation of CNT/MOSFET Based Active Electrode with ECG, EMG, and EEG Signals. (2020): 699 - 709.
MLA İstanbullu Mustafa,AVCI MUTLU Evaluation of CNT/MOSFET Based Active Electrode with ECG, EMG, and EEG Signals. , 2020, ss.699 - 709.
AMA İstanbullu M,AVCI M Evaluation of CNT/MOSFET Based Active Electrode with ECG, EMG, and EEG Signals. . 2020; 699 - 709.
Vancouver İstanbullu M,AVCI M Evaluation of CNT/MOSFET Based Active Electrode with ECG, EMG, and EEG Signals. . 2020; 699 - 709.
IEEE İstanbullu M,AVCI M "Evaluation of CNT/MOSFET Based Active Electrode with ECG, EMG, and EEG Signals." , ss.699 - 709, 2020.
ISNAD İstanbullu, Mustafa - AVCI, MUTLU. "Evaluation of CNT/MOSFET Based Active Electrode with ECG, EMG, and EEG Signals". (2020), 699-709.
APA İstanbullu M, AVCI M (2020). Evaluation of CNT/MOSFET Based Active Electrode with ECG, EMG, and EEG Signals. Çukurova Üniversitesi Mühendislik-Mimarlik Fakültesi Dergisi, 35(3), 699 - 709.
Chicago İstanbullu Mustafa,AVCI MUTLU Evaluation of CNT/MOSFET Based Active Electrode with ECG, EMG, and EEG Signals. Çukurova Üniversitesi Mühendislik-Mimarlik Fakültesi Dergisi 35, no.3 (2020): 699 - 709.
MLA İstanbullu Mustafa,AVCI MUTLU Evaluation of CNT/MOSFET Based Active Electrode with ECG, EMG, and EEG Signals. Çukurova Üniversitesi Mühendislik-Mimarlik Fakültesi Dergisi, vol.35, no.3, 2020, ss.699 - 709.
AMA İstanbullu M,AVCI M Evaluation of CNT/MOSFET Based Active Electrode with ECG, EMG, and EEG Signals. Çukurova Üniversitesi Mühendislik-Mimarlik Fakültesi Dergisi. 2020; 35(3): 699 - 709.
Vancouver İstanbullu M,AVCI M Evaluation of CNT/MOSFET Based Active Electrode with ECG, EMG, and EEG Signals. Çukurova Üniversitesi Mühendislik-Mimarlik Fakültesi Dergisi. 2020; 35(3): 699 - 709.
IEEE İstanbullu M,AVCI M "Evaluation of CNT/MOSFET Based Active Electrode with ECG, EMG, and EEG Signals." Çukurova Üniversitesi Mühendislik-Mimarlik Fakültesi Dergisi, 35, ss.699 - 709, 2020.
ISNAD İstanbullu, Mustafa - AVCI, MUTLU. "Evaluation of CNT/MOSFET Based Active Electrode with ECG, EMG, and EEG Signals". Çukurova Üniversitesi Mühendislik-Mimarlik Fakültesi Dergisi 35/3 (2020), 699-709.