Yıl: 2022 Cilt: 5 Sayı: 2 Sayfa Aralığı: 116 - 119 Metin Dili: İngilizce DOI: 10.14744/ijmb.2022.40316 İndeks Tarihi: 18-06-2022

Comparison of Diagon Coag XL and Cobas t 511 analyzers

Öz:
Objectives: We compared prothrombin time (PT), activated partial thromboplastin time (aPTT), fibrinogen, and Ddimer test results measured using the Diagon Coag XL coagulation analyzer with Cobas t 511 analyzer. Methods: Imprecision studies were performed for the PT, aPTT, fibrinogen, and D-dimer tests used by the Diagon Coag XL analyzer. For the comparison study, we used the leftover citrated plasma from patient samples after routine analysis with Cobas t 511. All of the results were analyzed using the correlation coefficient and Passing–Bablok regression analysis. Results: Total coefficient of variation obtained for all tests were within the criteria for acceptance. The method comparison study showed a good correlation between the results obtained on Diagon Coag XL and Cobas t 511 analyzers, except for aPTT test. The correlation coefficients obtained were 0.98 for PT, INR, and D-dimer, 0.95 for fibrinogen, and 0.80 for aPTT. Conclusion: For PT, aPTT, fibrinogen, and D-dimer tests, Diagon Coag XL analyzer is suitable for monitoring the coagulation system, and it can be used in clinical laboratories. However, the precision values of tests stated by the manufacturer must be verified.
Anahtar Kelime:

Belge Türü: Makale Makale Türü: Olgu Sunumu Erişim Türü: Erişime Açık
  • 1. Woodward M, Lowe GD, Rumley A, Tunstall-Pedoe H, Philippou H, Lane DA, et al. Epidemiology of coagulation factors, inhibitors and activation markers: The Third Glasgow MONICA Survey. II. Relationships to cardiovascular risk factors and prevalent cardiovascular disease. Br J Haematol 1997;97(4):785–97.
  • 2. Haas T, Fries D, Tanaka KA, Asmis L, Curry NS, Schöchl H. Usefulness of standard plasma coagulation tests in the management of perioperative coagulopathic bleeding: is there any evidence? Br J Anaesth 2015;114(2):217–24.
  • 3. Samuelson BT, Cuker A, Siegal DM, Crowther M, Garcia DA. Laboratory assessment of the anticoagulant activity of direct oral anticoagulants: a systematic review. Chest 2017;151(1):127– 38.
  • 4. Clinical and Laboratory Standards Institute (CLSI). Evaluation of precision performance of quantitative measurement methods; approved guideline 3rd ed. CLSI document EP05-A3. Wayne, PA: Clinical and Laboratory Standards Institute; 2014.
  • 5. Clinical and Laboratory Standards Institute (CLSI). One-stage prothrombin time (PT) test and activated partial thromboplastin time (APTT) test; approved guideline 2nd ed. CLSI document H47-A2. Wayne, PA: Clinical and Laboratory Standards Institute; 2008
  • 6. Clinical and Lab oratory Standards Institute (CLSI). Measurement procedure comparison and bias estimation using patient samples; approved guideline, 3rd ed. CLSI document EP09-A3. Wayne, PA: Clinical and Laboratory Standards Institute; 2013.
  • 7. Koç H, Yildiz Z, Hürmeydan Ö, Madenci ÖÇ, Yücel N, Orçun A. Evaluation of the analytical properties of the Diagon CoagXL coagulation analyzer. Clin Lab 2019;65(6).
  • 8. Can M, Guven B, Tekin A. Analytical performance assessment of the prothrombin time and activated partial thromboplastin time tests by Roche Cobas t511 Coagulation Analyzer. Indian J Biochem 2021;25(2):71–5.
  • 9. Lippi G, Salvagno GL, Gelati M, Poli G, Giavarina D, Favaloro EJ. Analytical assessment of the new Roche Cobas t 711 fully automated coagulation analyzer. Semin Thromb Hemost 2019;45(3):308–14.
  • 10. Park SJ, Chi HS, SH, Jang S, Park CJ. Evaluation of performance including influence by interfering substances of the Innovance D-dimer assay on the Sysmex coagulation analyzer. Ann Clin Lab Sci 2011;41(1):20–4.
  • 11. Mackie J, Lawrie AS, Kitchen S, Gaffney PJ, Howarth D, Lowe GD, et al. A performance evaluation of commercial fibrinogen reference preparations and assays for Clauss and PT-derived fibrinogen. Thromb Haemost 2002;87(6):997–1005.
APA Guven B, benice i, CAN M (2022). Comparison of Diagon Coag XL and Cobas t 511 analyzers. , 116 - 119. 10.14744/ijmb.2022.40316
Chicago Guven Berrak,benice ismail,CAN Murat Comparison of Diagon Coag XL and Cobas t 511 analyzers. (2022): 116 - 119. 10.14744/ijmb.2022.40316
MLA Guven Berrak,benice ismail,CAN Murat Comparison of Diagon Coag XL and Cobas t 511 analyzers. , 2022, ss.116 - 119. 10.14744/ijmb.2022.40316
AMA Guven B,benice i,CAN M Comparison of Diagon Coag XL and Cobas t 511 analyzers. . 2022; 116 - 119. 10.14744/ijmb.2022.40316
Vancouver Guven B,benice i,CAN M Comparison of Diagon Coag XL and Cobas t 511 analyzers. . 2022; 116 - 119. 10.14744/ijmb.2022.40316
IEEE Guven B,benice i,CAN M "Comparison of Diagon Coag XL and Cobas t 511 analyzers." , ss.116 - 119, 2022. 10.14744/ijmb.2022.40316
ISNAD Guven, Berrak vd. "Comparison of Diagon Coag XL and Cobas t 511 analyzers". (2022), 116-119. https://doi.org/10.14744/ijmb.2022.40316
APA Guven B, benice i, CAN M (2022). Comparison of Diagon Coag XL and Cobas t 511 analyzers. International Journal of Medical Biochemistry, 5(2), 116 - 119. 10.14744/ijmb.2022.40316
Chicago Guven Berrak,benice ismail,CAN Murat Comparison of Diagon Coag XL and Cobas t 511 analyzers. International Journal of Medical Biochemistry 5, no.2 (2022): 116 - 119. 10.14744/ijmb.2022.40316
MLA Guven Berrak,benice ismail,CAN Murat Comparison of Diagon Coag XL and Cobas t 511 analyzers. International Journal of Medical Biochemistry, vol.5, no.2, 2022, ss.116 - 119. 10.14744/ijmb.2022.40316
AMA Guven B,benice i,CAN M Comparison of Diagon Coag XL and Cobas t 511 analyzers. International Journal of Medical Biochemistry. 2022; 5(2): 116 - 119. 10.14744/ijmb.2022.40316
Vancouver Guven B,benice i,CAN M Comparison of Diagon Coag XL and Cobas t 511 analyzers. International Journal of Medical Biochemistry. 2022; 5(2): 116 - 119. 10.14744/ijmb.2022.40316
IEEE Guven B,benice i,CAN M "Comparison of Diagon Coag XL and Cobas t 511 analyzers." International Journal of Medical Biochemistry, 5, ss.116 - 119, 2022. 10.14744/ijmb.2022.40316
ISNAD Guven, Berrak vd. "Comparison of Diagon Coag XL and Cobas t 511 analyzers". International Journal of Medical Biochemistry 5/2 (2022), 116-119. https://doi.org/10.14744/ijmb.2022.40316